The Fluke 2635T
Recording Thermometer is a precision multi channel
temperature recording and logging instrument that
delivers up to 0.1 degree accuracy (3 sigma, one year
accuracy specification) for temperature monitoring and
calibration applications. Based on Fluke's popular 20
channel, new Hydra Series II data loggers, these units
are matched with a precision SPRT probe (RTD) on channel
one and calibrated as a system for maximum accuracy and
precision. The unit is delivered with current
calibration data, ready for use. This recording
thermometer includes a NIST traceable and DKD certified
calibration (with certificate) that covers the Hydra
Series II and probe as a system. A soft protective case
is supplied to protect the probe when not in use. Hydra
Logger Software is also included for support of basic
data collection and easy instrument configuration.
Optional Trend Link for Fluke trending software (model
2600A-904) is also available for electronic chart
recording display and analysis. In addition to being a
precision thermometer, it is capable of measuring up to
18 additional channels of thermocouple, dc and ac Volts,
frequency, and 2-wire Ohms. A total of 10 RTD probes can
be connected and recorded at one time.
The 2635T is based on a Fluke
Hydra Series II Data Bucket Data Logger which uses a
PCMCIA memory card (256K) for storage of time stamped
data. Data files may be imported to any standard spread
sheet for analysis and documentation. The 2635A user's
manual (PN 932160) reflects all operational
specifications and unit information except the 2635T's
special calibration data, which is specific to each
unit.
Probe hook up diagram:
The probe connects to channel
1(input) and 11(excitation) on the Hydra input
connector.

|