| Automated AC
measurement with precision that is easy to use
The 5790A is a complete
automated ac measurement standard designed for the most
demanding calibration applications. It combines the
accuracy you would expect from a thermal transfer
standard with the ease-of-use of a digital multimeter.
Absolute ac voltage measurement uncertainties are as low
as ±24 ppm (one year, 23oC ±5oC). The 5790A is
designed to meet the complete ac voltage and wideband
verification requirements of the Fluke 5700A, 5500A and
other calibrators, amplifiers like the 5725A and 5205A,
and transfer standards and ac voltmeters.
The 5790A covers an
alternating voltage range of 700 µV to 1000V, and a
frequency range of 10 Hz to 1 MHz. A wideband voltage
option extends frequency range to 30 MHz to meet the
calibration requirements of Fluke 5700A, 5720A and 5100
Series Wideband Options.
The 5790A is also
compatible with Fluke A40 and A40A Current Shunts, which
permit you to make ac/dc current transfer measurements
up to 20A.
The 5790A may be used
alone or as a transfer standard with an external dc
source. In either case the normally tedious switching
and calculations are performed automatically by the
5790A, and the resulting ac/dc difference is displayed
directly on the easy-to-read vacuum fluorescent display.

|